VentilCare
Low cost ventilators that can be used in the emergency cases amid covid19.
In amid COVID-19 hospitals in india might face availability of approved ventilators. We are building low cost ventilator by automating BVM device. Which is expected to cost less than 7k but further cost would depend on the 3D printing material.
Right now We are developing with the available components that we have..
Our approach to the project:
We are targeting non-invasive ventilator here. There are two challenges,
1. Procurement of medical device(BVM)
2. Automating the BVM device and procuring the same.
why it is needed ?
According to the study There are about 70000 icu beds and 40000 ventilators are present in total in india. https://theprint.in/opinion/current-rate-india-30000-covid-19-deaths-may-no-hospital-bed-june-data/385386/
since covid 19 cases are exponentially rising in india, this would cause in the scarcity in the availability of approved ventilators which might later cost 30000 deaths.
there are atleast 30000 no of ventilators would essentially be needed in the country. Hence locally made low cost tested ventilators should be made available immediately.
After going through the specifications set by MHRA. following learnings are made.
a. Must have mandatory ventilation.
b. Inspiratory:Expiratory ratio (I:E). The proportion of each breathing cycle that is spent breathing in compared to breathing out. a. 1:2.0 (i.e. expiration lasts twice as long as inspiration). Optionally adjustable in the range 1:1–1:3.
c. Respiratory Rate. The number of breathing cycles every minute. Range 10–30 breaths per minute in increments of 2 (only in mandatory mode) can be set by the user.
d. The ventilator needs to be closed loop to prevent microbes from spreading in the room and infecting other patients.
E. The ventilator needs to be non-invasive.
F. Positive End Expiratory Pressure (PEEP). The pressure maintained in the breathing system during expiration. a. Range 5–20 cm H2O adjustable in 5 cmH2O increments. b. Patient breathing system must remain pressurised to at least the PEEP level setting at all times.
Key Ventilation Specifications for Designing the mechanical part.
- Note changes to specifications with increased BPM and I:E ratios. Adjusting one of these parameters may influence the other, since at higher BPM, the tidal volume is usually decreased. Consult a clinician.
- From the Clinical Guidance this document summarizes the minimum set of requirements for ventilation:
- Patients must be under the management of a trained clinician.
- The minimum controllable parameters in order to ventilate a patient include:
- Respiratory Rate (RR) (breaths per minute): between 8–40
- Tidal Volume (TV) (air volume pushed into lung): between 200–800 mL based on patient weight
- I/E Ratio (inspiratory/expiration time ratio): recommended to start around 1:2; best if adjustable between range of 1:1–1:4*
- Assist Detection pressure. When a patient tries to inspire, they can cause a dip on the order of 1–5 cm H2O, with respect to PEEP pressure (not necessarily = atmospheric).
- Airway pressure must be monitored
- Maximum pressure should be limited to 40 cm H2O at any time; Plateau pressure should be limited to max 30 cm H2O
- The use of a passive mechanical blow-off valve fixed at 40 cm H2O is strongly recommended
- Clinician require readings of plateau pressure and PEEP (refer to clinical documentation tab)
- PEEP of 5–15 cm H2O required; many patients need 10–15 cmH2O
- Failure conditions must permit conversion to manual clinician override, i.e. if automatic ventilation fails, the conversion to immediate ventilation must be immediate.
- Ventilation on room air is better than no ventilation at all. Blending of oxygen and air gas mixture to adjust FiO2 is not important in an emergency scenario. It is certainly nice to have that ability and can easily be implemented with an oxygen / air gas blender that some hospitals already have.
- COVID-19 can get aerosolized (airborne), so HEPA filtration on the patient’s exhalation is required or between the ventilator unit and the patient (at the end of the endotracheal tube) to protect clinical staff from certain infection. In-line HEPA filters can usually be purchased alongside manual resuscitator bags.
- Heat and moisture exchanger should be used in line with the breathing circuit.
- Failure conditions must result in an alarm.
This is a minimum requirement set for emergency use. Equipment designed for more regular use, even if for emerging markets, will require additional features to be used on a regular basis.
*Range determined based on several COVID-19 patients’ ventilator settings reported from Boston area ICUs
Get Arun Kothinti’s stories in your inbox
Join Medium for free to get updates from this writer.
Things to do before building it.
- Develop compatible algorithms to align blower cycle to the breathing cycle by looking at back pressure load on the blower, or very simple air pressure sensor that can be inserted into the output of the pump. For ease of developing algorithms see Colab notebook with data visualization.
- Solutions to prevent the system from becoming unsafe if power is lost. It’s not obvious to me on how to do this. Ideas are welcome.
- Test with a simple inflator pump and achieve the same pressure control.
- Develop a low-cost simple minute volume sensor/approximator. Here is some information about the downsides of being over/under on target volume. Critical care seems to want up to 10–12L/Min.
Approach and Materials Required
Automating the BVM: I’ll explain in brief below why we are doing this,
Covid-19 infected patients will highly suffer from breathing difficulties, respiratory distress and also to avoid the spread they will isolate the patient respiratory track using an external ventilator. This ventilator provides 60% to 80% of oxygen content based on the severity of patients which get pumped directly into the lungs. This will maintain the oxygen content in blood within the normal range. The existing ventilator are the advanced version where most of the features are not utilized at full extinct. For the extreme critical survival an already existing mechanical ventilator are available. These ventillator mostly will get used in CPR and at critical stages. By automating this mechanical ventillator based on the medically approved specification on respiratory rate. This model can be built.
Approach:
The design model should be accurate in pressure spec where it get pumped through and fro into the respiratory track.
Strict FDA guidelines should be followed while designing this model.
Respiratory rate:
This should setted manually by doctors after examining the patients.
Inhale will be re-configurable at any given time and also options will be given for standard value. (There wont be any control on exhale)
Linear actuators will be selected for pumping the ambu bag with casing upfront.
Add-On:
1. We can include the oximeter to monitor the oxygen content in blood
2. ECG sensor to stream the pump level
3. We can integrate this using wifi enabled devices to monitor huge patients data.
4. Patient critical alert system (Auto & Manual).
- a powerful microcontroller Compatible Nano/Clone to control the motor speed, breathing cycles, and handle user input.
- Brushless DC Motor Electronic Speed Controller capable of at least 2A.
- Button or switch for input control. The code cycles through 5 levels of pressure.
- 12v DC power supply that can handle at least 5A. The blower works pretty hard, and the breathing cycles can create voltage drops which will reset the Arduino if the supply is not strong enough.
- Tubing and face mask.We use 3D printable parts.
- Optional inline 12v battery backup.
- Air pressure sensor
- FTDI USB to serial Adapter
- voltage adapter, dotted PCB,OPTOISOLATOR
Building the Ventilcare
The required designs and the pre production details will be posted time to time.
DISCLAIMER:
DO NOT implement this system without design inputs from appropriate medical professionals.
- DO NOT implement this system without the permission and involvement of appropriate medical professionals.
- ENSURE hygiene and consistency of performance before providing support to a patient.
- This design has not been tested aggressively on patients and is my DIY thought approach to building low cost ventilators for situations of emergency.
- I do not take responsibility for any mishaps/casualties that may occur due to change in design/inappropriate installation.
THIS IS MADE OPEN SOURCE FOR LEARNING.
COMMERCIAL USE IS STRICTLY PROHIBITED.
NOTE:
YOU CAN ALSO CONTRIBUTE TO THIS PROJECT WITH NECESSARY CHANGES FOR THE BETTER EFFICIENCY.